| Measurement Range: | (0.00 - 99.99)% (intermediate ranges are selectable). |
| Maximum Allowable Error: | ±2%FS. |
| Resolution: | 0.01%. |
| Repeatability: | ±1%FS. |
| Warm - up Time: | ≤15 min. |
| Response Time: | T90 ≤ 15 s. |
| Company Info. |
| Xi'an Kacise Optronics Co.,Ltd. |
| Verified Supplier |
| View Contact Details |
| Product List |
KTG120 Hydrogen Analyzer (3U Online Type) for Petrochemical Industry
Introduction
The KTG120 Hydrogen Analyzer is an online detection instrument for the hydrogen content in mixed gases. It is developed by combining imported micro - flow thermal conductivity gas sensors with the latest microcomputer processing technology.
Application Areas
It is used for the non - explosion - proof online analysis of the hydrogen content in mixed gases in industries such as petrochemical, metallurgy, power, and air separation, or for the non - explosion - proof online analysis of the hydrogen content in hydrogen production from ammonia decomposition.
Instrument Features
Technical Parameters
Measurement Range: (0.00 - 99.99)% (intermediate ranges are selectable).
Maximum Allowable Error: ±2%FS.
Resolution: 0.01%.
Stability: Zero drift: ±2%FS/7d; Span drift: ±2%FS/7d.
Repeatability: ±1%FS.
Warm - up Time: ≤15 min.
Response Time: T90 ≤ 15 s.
Sample Gas Flow Rate: (300 ± 10) mL/min.
Sample Gas Pressure: 0.05 MPa ≤ Inlet pressure ≤ 0.25 MPa.
Analog Output: (4 - 20) mA standard signal.
Digital Output: A standard RS485 communication port enables two - way communication with a computer, and an Ethernet port enables connection to the wide - area network.
Contact Output: Dual - relay output: 220VAC, 1A or 24VDC, 2A.
Working Environment: Temperature: - 10℃ - + 45℃; Humidity: ≤90%RH.
Working Power Supply: (220 ± 22) VAC, (50 ± 5) Hz.
External Dimensions: 480mm (width) × 136mm (height) × 370mm (depth).
Installation Dimensions: 442mm (width) × 134mm (height).
Weight: Approximately 3.3 Kg.
|