Refer to Fiatest TM Fluorescence Immunoassay Analyzer Operation Manual for the
complete instructions on use of the Test. The test should be in
room temperature.Allow the test cassette, specimen, buffer and/or
controls to reach room temperature (15-30°C) prior to testing.
1. Turn on the Analyzer power. Then according to the need, select
“standard test” or “Quick test” mode.
2. Remove the ID card and insert it into the Analyzer port.
3. Pipette 10μL of specimen into the buffer tube, then mix
thoroughly.
4. Pipette 75μL diluted sample into the sample well. Start the
timer at the same time.
5. There are two test modes for Fiatest TM Fluorescence Immunoassay Analyzer,Standard Test mode and Quick
Test mode. Please refer to the user manual of Fiatest TM Fluorescence Immunoassay Analyzer for details.
“Quick test” mode: After 10 minutes of adding sample,Insert the
test cassette into the Analyzer and click "New test"
immediately,The Analyzer will automatically give the test result
after a few seconds.
“Standard test” mode: Insert the test cassette into the Analyzer
immediately after adding specimen, click "New test" at the same
time, The Analyzer will automatically countdown 10 minutes. After
the countdown, the Analyzer will give the result at once.